After pulling the completed Flongle run of Group 1, Library 1, performed pore check on all remaining Flongle flow cells.
As a reminder, pore check of the flow cell used for Group 1 Library 1 showed:
SN | Flow Cell ID | Num. pores |
---|---|---|
B028104085 | AYW935 | 59 |
Checked all 11 remaining Flongle flow cells. As a reminder, <50 pores would fall below warranty, and the maximum is 126pores.
SN | Flow Cell ID | Num. pores |
---|---|---|
B028104112 | AXS867 | 37 |
B028103996 | AXS408 | 40 |
B028103978 | AYL256 | 35 |
B028103949 | AYL126 | 80 |
B028104129 | AYW871 | 61 |
B028104068 | AYK267 | 59 |
B028103992 | AYK963 | 89 |
B028104122 | AYW685 | 59 |
B028104029 | AYX067 | 79 |
B028104142 | AYL029 | 44 |
B028104167 | AYL248 | 60 |
Wow, so 4/12 (or 1/3) of the Flongle flow cells we received are below warranty. Another 5, including the cell used for Group 1 Library 1, have pore counts just above the minimum.
I contacted Nanopore Support to request replacements for the out-of-warranty cells on 08/13/2025, case number 01333308. Reorder number is RD060692.